End-station 1:
The STOE single-crystal diffractometer on End-station 1 is currently operating with a RDI 1M CMOS detector.
Some specs here
End-Station 2:
End-station 2 currently employs 2 area detectors:
Mar345 image plate reader:
Pixel size: Variable: 100 um x 100 um or 150 um x 150 um
Dynamic range: 16 bit, no readout noise
Active Area: circular with 345 mm diameter
Closest distance to diffraction volume: ~ 270 mm
Cycle time: ~ 90 secs read out plus exposure time; needs ~ 90 secs erasure with X-ray alignment activities between exposure
Exposure time is shutter controlled via LabView link
File format: .mar345 format (readable by most diffraction data reduction software (Dioptas, fit2d)
File size: ~ 8 MB (100 um pixel size), ~ 5 MB (150 um pixel size)
Dectris: Pilatus3 S 1M Hybrid Photon Counting detector
Pixel size: 172 um x 172 um
Dynamic range: 20 bit, no readout noise
Active Area: rectangular 168.7 (h) x 179.4 (v) mm (981 x 1043 pixels). This includes 1 vertical (7 pixels) and 4 horizontal (17 pixels) gaps.
Closest distance to diffraction volume: ~ 250 mm
Cycle time: < 100 msec readout, i.e. controlled by exposure time
Exposure time is electronically gated. No erasure cycle needed after X-ray activities.
File format: tiff or cbf
File size: ~ 4 MB
- More info here